An instrument to detect atmospheric concentrations of the hydroxyl (OH) and hydroperoxyl (HO 2 ) radicals has been developed using the FAGE (fluorescence assay by gas expansion) technique. The instrument is housed in a mobile laboratory and monitors the OH radical via on-resonance laser-induced fluorescence (LIF) spectroscopy of the A 2 Σ + (v′=0)–X 2 Π i (v ″ =0) transition at ca. 308 nm. Ambient air is expanded through a 1 mm nozzle to low pressure where it is irradiated by the laser pulse at a repetition rate of 7 kHz, with the resultant fluorescence being detected by gated photon counting. HO 2 is monitored by chemical conversion to OH by the addition of NO, with subsequent detection using LIF. Following laboratory and field calibrations to characterise the instrument sensitivity, detection limits of 1.8×10 6 and 2.1×10 7 molecule cm -3 were determined for OH and HO 2 respectively, for a signal-to-noise ratio, S/N, of 1 with 150 s signal integration time. The instrument was deployed for the first time during the ACSOE field campaign at Mace Head, Eire, for which illustrative results are given.
No takes yet. Share an insight, caveat, or question.
Creasey et al. (1997) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: