By utilizing a reflection geometry, an accurate mass absorption coefficient of a sample can be determined by measuring the Ag Kα Compton intensity. Intensities of Ag Kα Compton scattered radiation have been collected by using either the usual reflection geometry of a Philips PW 1450 automatic X‐ray spectrometer or a more refined reflection geometry, achieved on a Philips PW 1540/10A manual X‐ray spectrometer. The experimental results have shown that the relationship between the Ag Kα Compton intensity and the mass absorption is a logarithmic function. Our experimental results are not in agreement with those reported in literature, but a theoretical explanation to account for this fact has not been achieved as yet.
No takes yet. Share an insight, caveat, or question.
Franzini et al. (1976) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: