We present data demonstrating the influence of an applied electric field E oriented normal to the input surface of a microchannel plate (MCP) detector on the critical operating parameters of the detector, including the quantum detection efficiency, the spatial resolution, and pulse height distribution. The MCP detector response is characterized using 20 keV protons as the primary radiation. An applied electric field E<−4 V/mm, where a negative value of E corresponds to a nearby object that is biased positive relative to the input surface, results in a high spatial resolution and a quantum detection efficiency that is approximately equal to the open area ratio of the MCP. An electric field −1
No takes yet. Share an insight, caveat, or question.
Funsten et al. (1996) studied this question.