We performed an absolute frequency measurement of an acetylene stabilized laser utilizing a femtosecond injection locking technique that can select one component among the fiber laser comb modes. The injection locking scheme has all the fiber configurations. Femtosecond comb lines of 250 MHz spacing based on the fiber femtosecond laser were used for injection locking of a distributed feedback (DFB) laser operating at 1542 nm as a frequency reference. The comb injected DFB laser serves as a selection filter of optical comb modes and an amplifier for amplification of the selected mode. The DFB laser injection locked to the desired comb mode was used to evaluate the frequency stability and absolute frequency measurement of an acetylene stabilized laser. The frequency stability of the acetylene stabilized laser was measured to be 1.1 x 10(-12) for a 1 s averaging time, improving to 6.9 x 10(-14) after 512 s. The absolute frequency of the laser stabilized on the P(16) transition of (13)C(2)H(2) was measured to be 194 369 569 385.7 kHz.
No takes yet. Share an insight, caveat, or question.
Ryu et al. (2008) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: