NIRS measurements of tissue oxygen saturation show increasing within-subject variability at higher exercise workloads and should be interpreted with caution.
The present study evaluated the reproducibility of tissue oxygenation in relation to oxygen consumption (VO2) across cycle exercise intensities in a test-retest design. 12 subjects (25.7±2.1 years; 24.7±1.9 kg · m(-2)) twice performed an incremental bicycle exercise protocol, while tissue oxygen saturation (StO2) in the vastus lateralis muscle was monitored by a commercially available NIRS unit and VO2 determined by an open-circuit indirect calorimetric system. Coefficients of variation across rest, workloads corresponding to 25, 50 and 75% of individual maximum capacity, and maximum load were 5.8, 4.6, 6.1, 8.0, 11.0% (StO2) and 7.6, 6.0, 3.7, 3.4, 3.1% (VO2), respectively. 95 % CI of relative test-retest differences ranged from -5.6 to +5.4% at 25% load to -17.2 to +7.5% at maximum load for StO2 and from -7.3 to +7.7% at rest to -3.3 to +3.2% at maximum load for VO2. With advancing exercise intensity, within-subject variability of StO2 was augmented, whereas VO2 variability slightly attenuated. NIRS measurements at higher workloads need to be interpreted with caution.
No takes yet. Share an insight, caveat, or question.
Thiel et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: