Laboratory evaluation demonstrates high arsenic removal from synthetic water using reverse osmosis filters, highlighting effective domestic mitigation of toxic exposures.
Arsenic (As) in drinking water is known to cause serious health issues, and point-of-use under-sink filtration systems are commonly used to mitigate arsenic exposure at the domestic level. The arsenic removal performance of a commercially available reverse osmosis (RO)-based under-sink filtration system was assessed using synthetic drinking water samples and inductively coupled plasma mass spectrometry (ICP-MS). Arsenic-contaminated synthetic drinking water samples were filtered in the laboratory and analyzed for arsenic content using ICP-MS. The accuracy of the analytical procedure was confirmed by analyzing certified reference materials, namely Enviromat Wastewater Low (EU-L), NIST SRM 1643f, and UME CRM 1201. The obtained recovery percentages ranged from 97.28% to 100.5%, confirming the accuracy and reliability of the analytical technique. The arsenic removal efficiencies achieved by the filtration system ranged from 97.41% to 98.96%, while effluent concentrations remained below internationally accepted drinking water guideline values. The method detection and quantification limits were found to be 0.028 µg/L (LOD) and 0.095 µg/L (LOQ), respectively, indicating sufficient sensitivity for trace-level arsenic measurement.