Recently manufactured intraocular lenses (IOLs) (STAAR AA 4203 made of silicone and IOGEL PC-12 made of polyHEMA) were tested before and after implantation simulation for axial optical resolution (model eye), for optical surface quality (photokeratoscope), and for surface microtrauma (scanning electron microscopy). At a 3 mm pupillary opening, both lens types achieved a mean axial resolution of 1.25 (range 1.0 to 1.5) on an optotype analogous resolution target. When the pupillary opening was increased to 6 mm, polymacon IOLs showed only a moderate resolution decrease (up to one line) and loss of contrast (up to 30%); silicone IOLs showed almost no decrease. Photokeratoscopic evaluation revealed that a circular optical nonhomogeneity at the peripheral optical zone of the hydrogel IOLs is the source of this impairment. Groove-like surface deformities occasionally observed by scanning electron microscopy did not detectably decrease resolution or contrast.
No takes yet. Share an insight, caveat, or question.
Kulnig et al. (1990) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: