Digital Fourier filtering is used to produce a temperature-insensitive univariate calibration model for measuring lysozyme in aqueous solutions. Absorbance spectra over the 5000-4000 cm-1 spectral range are collected for lysozyme standards maintained at 14 degrees C. These spectra are used to compute the calibration model while a set of spectra collected at temperatures ranging from 4 to 24 degrees C are used to validate the accuracy of this model. The root-mean-square error of prediction (RMSEP) is 0.279 mg/mL over a tested lysozyme concentration range of 0.036-51.6 mg/mL. The detection limit is 0.68 mg/mL. In addition, multivariate calibration models based on partial least-squares regression (PLS) are evaluated and compared to the results from the univariate model. PLS outperforms the univariate model by providing a RMSEP of 0.090 mg/mL. Analysis of variance showed that both calibration methods effectively eliminate the adverse affects created by variations in solution temperature.