A low-field medium-resolution NMR spectrometer, with an operating frequency of 29 MHz for 1 H, has been developed for use in process analysis. The information that is attainable at this field strength has been investigated with the use of samples from a wide range of applications including the qualitative monitoring of a benzene production process, the estimation of the average ethoxy chain length in nonyl phenol ethoxylates, and the determination of strong acid concentrations. At low field, the spectra of multi-component samples often exhibit overlapping peaks, so multivariate calibration methods are necessary for quantitative analysis. Simulated data have been used to illustrate the effects of signal-to-noise ratio (SNR), resolution, and line shape on the predictive ability of partial least-squares (PLS) calibration routines, with SNR shown to have a greater impact than resolution on accuracy and precision. PLS calibration models were applied successfully in the analysis of samples from a methacrylamide process and a simulated esterification reaction. The accuracy (<5%) and precision (<3%) were generally good for components at concentrations >1–5% w/w.