The depth distribution of the thermal neutron flux is a major factor in boron neutron capture therapy (BNCT) in determining the efficiency of cell sterilization. In this paper the fission detector method is developed and applied to measure the in-phantom thermal neutron flux depth distribution. Advantages of the fission detector include small size, direct measurement of thermal neutron flux in a mixed radiation field of BNCT beam, self-calibration, and the possibility of on-line measurement. The measurements were performed at epithermal a BNCT facility. The experimental results were compared with the thermal neutron flux calculated by the Monte Carlo method and found to be in good agreement.
No takes yet. Share an insight, caveat, or question.
Kaplan et al. (1999) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: