An improved Dual-wavelength-excitation Particle Fluorescence Spectrometer (DPFS) has been reported. It measures two fluorescence spectra excited sequentially by lasers at 263 nm and 351 nm, from single atmospheric aerosol particles in the 1-10 mum diameter size range. Here we investigate the different levels of discrimination capability obtained when different numbers of excitation and fluorescence-emission wavelengths are used for analysis. We a) use the DPFS to measure fluorescence spectra of Bacillus subtilis and other aerosol particles, and a 25-hour sample of atmospheric aerosol at an urban site in Maryland, USA; b) analyze the data using six different algorithms that employ different levels of detail of the measured data; and c) show that when more of the data measured by the DPFS is used, the ability to discriminate among particle types is significantly increased.
No takes yet. Share an insight, caveat, or question.
Pan et al. (2010) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: