Answers are grounded in this paper and cite their sources.
This study introduces a refined current feedback instrumentation amplifier (CFIA) specifically designed for amplifying biopotential signals that originate from resistive-bridge sensors. The proposed architecture uniquely incorporates a multipath chopper-stabilized CFIA which has been developed to minimize the effects of bridge offsets, while maintaining low power usage, high input impedance, and reduced noise characteristics. The engineering blueprint employs a ripple reduction loop (RRL) to mitigate output ripple caused by chopper up-modulation. To speed up offset cancellation and to limit the offset caused by bridge mismatch, an automatic offset cancellation loop (AOCL) circuit is integrated within the analog front end (AFE). Crafted using a conventional 0.18 µm CMOS process, the CFIA in this design provides adjustable gain between 20.35 dB to 55.14 dB, with a power supply rejection ratio (PSRR) and a common mode rejection ratio (CMRR) of 103 dB and 114 dB, respectively. The system demonstrates an input-referred noise (IRN) value of 16 ηV/√Hz at 200 Hz frequency, equivalent to a noise efficiency factor (NEF) of 5.18. Operating from a supply voltage of 1.8V, the AFE shows a power consumption of 291 μW.
No takes yet. Share an insight, caveat, or question.
Laababid et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: