August 29, 2026Annals of the ICRP

Development of a pragmatic methodology for measurement of dose for KV CBCT in radiotherapy: A pilot study

View Full Paper
Ask AI
Bookmark
Share

Authors

MDM. DjukelicSGS. GrosTKT. Kron

Discussion

Loading...

Member takes

Overview

Comparative study demonstrates standard clinic equipment accurately measures cone-beam computed tomography doses, suggesting radiation oncology centres can readily optimize imaging protocols.

Key Points

  • To develop and evaluate a practical dose measurement methodology for kilovoltage cone beam computed tomography using equipment readily available in standard radiotherapy clinics.
  • Assessed an alternative measurement arrangement comprising a 0.6 cc Farmer-type ionization chamber inside a 30 cm × 30 cm × 30 cm cubic phantom made of water-equivalent slabs.
  • Measured weighted dose index values from linear accelerator-based kilovoltage cone beam computed tomography and compared them against measurements from standard 100 mm chambers in cylindrical PMMA phantoms.
  • Weighted dose index values obtained with the alternative setup were determined to be a viable surrogate for conventional cylindrical CT phantom measurements.
  • The methodology provides an accessible approach for radiotherapy centers lacking dedicated CT dosimetry phantoms to quantify radiation dose and optimize imaging protocols.