Key result
Restenosis caused by drug-eluting stent fracture in the vertebral artery origin was successfully managed in two patients using the stent-in-stent technique.
Case Report (n=2)
This first report of drug-eluting stent fracture in the vertebral artery origin demonstrates that resulting restenosis can be successfully managed using a stent-in-stent technique.
Answers are grounded in this paper and cite their sources.
May offer bailout option for vertebral artery stent fracture restenosis; case report leaves open durability and broader applicability.
The authors report two cases of stent fracture and restenosis after placement of a drug-eluting device in the vertebral artery (VA) origin, and describe management of restenosis with the stent-in-stent technique. Two women, one 62 and the other 67 years of age, underwent stent placement in the VA origin to treat symptomatic and angiographically significant stenosis in this vessel. Sirolimus-eluting coronary stents (Cypher) were used in both cases. Four months after placement of the devices, the symptoms recurred. Follow-up angiography performed 5 months after insertion of the devices revealed a transverse stent fracture with separation of the fragments and severe in-stent restenosis in both cases. The restenoses were treated with reinsertion of coronary stents (Cypher and Jostent FlexMaster) by using the stent-in-stent technique. After stent reinsertion, the patients exhibited relief of symptoms. This paper is the first report of fracture in a drug-eluting stent and restenosis after stent placement in the VA origin. Restenosis caused by such a fracture can be managed successfully by performing the stent-in-stent maneuver. The physical properties of metallic devices, stent strut geometry, and anatomical peculiarities of the subclavian artery may be associated with stent fractures. Earlier follow-up angiography studies (within 6 months) are warranted.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2007) conducted a case report in Vertebral artery origin stenosis (n=2). Stent-in-stent technique was evaluated. Restenosis caused by drug-eluting stent fracture in the vertebral artery origin was successfully managed in two patients using the stent-in-stent technique.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: