Key result
Selective Systolic Contouring at the aortic valve level yielded the smallest bias in estimating aortic forward flow compared to standard contouring methods (-5 ml vs -11 ml, p<0.001).
Why the study?
Optimizing MRI aortic flow quantification is crucial for accurate assessment of valvular disease severity, prompting evaluation of a novel contouring method versus standard methods.
Does Selective Systolic Contouring (SSC) improve the accuracy of MRI aortic flow quantification compared to standard contouring in patients with aortic valve disease?
Observational (n=98)
No
Does Selective Systolic Contouring (SSC) improve the accuracy of MRI aortic flow quantification compared to standard contouring in patients with aortic valve disease?
Selective Systolic Contouring (SSC) at the aortic valve level provides a more accurate MRI assessment of aortic forward flow than standard contouring, which may mitigate errors in estimating mitral regurgitation.
Optimizing MRI aortic flow quantification is crucial for accurate assessment of valvular disease severity. In this study, we sought to evaluate the accuracy of a novel method of contouring systolic aortic forward flow in comparison to standard contouring methods at various aortic levels. The study included a cohort of patients with native aortic valve (AoV) disease and a small control group referred to cardiac MRI over a 1-year period. Inclusion criteria included aortic flow quantification at aortic valve and one additional level, and no or trace mitral regurgitation (MR) documented both by the MRI AND an echocardiogram done within a year. In addition to flow quantification with standard contouring (SC), a novel Selective Systolic Contouring (SSC) method was performed at aortic valve level, contouring the area demarcated by the AoV leaflets in systole. The bias in each technique's estimate of aortic forward flow was calculated as the mean difference between aortic forward flow and left ventricular stroke volume (LV SV). 98 patients (mean age 56, 71% male) were included: 33 with tricuspid and 65 with congenitally abnormal (bicuspid or unicuspid) AoV. All methods tended to underestimate aortic forward flow, but the bias was smallest with the SSC method (p < 0.001). Therefore, SSC yielded the lowest estimates of mitral regurgitant volume (4.8 ml) and regurgitant fraction (3.9%) (p < 0.05). SSC at AoV level better approximates LV SV in our cohort, and may provide more accurate quantitative assessment of both aortic and mitral valve function.
No takes yet. Share an insight, caveat, or question.
Yaman et al. (2024) conducted an observational in Aortic valve disease (n=98). Selective Systolic Contouring (SSC) vs. Standard contouring (SC) was evaluated on Bias in estimate of aortic forward flow (mean difference between aortic forward flow and left ventricular stroke volume) (p=<0.001). Selective Systolic Contouring at the aortic valve level yielded the smallest bias in estimating aortic forward flow compared to standard contouring methods (-5 ml vs -11 ml, p<0.001).