An amperometric biosensor immobilizing two enzymes and an electron mediator in an identical plane has been fabricated by the self-assembly technique for determination of methanol in crude plant samples. A self-assembled mixed monolayer of 4,4'-dithiodibutyric acid covalently attached two enzymes (Hansenula sp. alcohol oxidase and horseradish peroxidase) and 11-ferrocenyl-1-undecanethiol as an electron mediator on an Au electrode is exploited to produce a two-dimensional reaction matrix. The composition of the two enzymes and electron mediator molecules was optimized for detection of methanol in 0.1 M sodium phosphate buffer (pH 6.0). We successfully quantified methanol in low-purity tobacco (Nicotiana tabacum) plant extracts with the biosensor, which showed sensitivity comparable to that of gas chromatography/mass spectrometry. The redox-relay biosensor is quite simple and stable due to its covalent attachment to the Au surface, making it possible to downsize the construction. We fabricated a miniature methanol biosensor that fitted a well of a 96-well micro assay plate available for high-throughput assay. The biosensor is advantageous for the sensitive, continuous, and convenient determination of methanol.
No takes yet. Share an insight, caveat, or question.
Hasunuma et al. (2004) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: