PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
October 8, 2025Magnetic Resonance in Medicine0 citations

Impact of Lower Limb Amputation on Implant Heating During 1. 5 T MRI : Assessment of Active and Passive Device

View Full Paper
Ask AI
Bookmark
Share

Why the study?

The study investigated the impact of lower limb amputation on radiofrequency-induced heating of active and passive implantable medical devices in a 1.5 T MRI environment.

Does lower limb amputation impact RF-induced heating of active and passive implantable medical devices during 1.5 T MRI?

Population

High-resolution anatomical models and an anatomically inspired leg phantom with embedded metallic implants

Comparison

Levels of lower limb amputation (below-knee vs above-knee)

Design

Electromagnetic simulation and experimental phantom validation study

Key result

More extensive lower limb amputations reduced overall specific absorption rate and temperature rise for implants during 1.5 T MRI, but intensified localized heating near the amputation site.

Authors

ASAo ShenMAMir Khadiza AkterZZZiyu Zuo

Discussion

Loading...

Member takes

Overview

Localized heating risks warrant caution for MRI in amputees with implants; leaves open human translation from animal data.

Structured PICO

Does lower limb amputation impact RF-induced heating of active and passive implantable medical devices during 1.5 T MRI?

P
Population
High-resolution anatomical models representing various body types and levels of amputation (below-knee and above-knee) and an anatomically inspired leg phantom with embedded metallic implants
E
Exposure
1.5 T MRI environment
C
Comparator
Different levels of amputation (below-knee vs above-knee)
O
Outcome
Specific absorption rate (SAR) near modular hip prostheses and RF-induced lead-tip heating for a peripheral nerve stimulation (PNS) devicesafety

Lower limb amputation alters RF field distribution during 1.5 T MRI, necessitating amputation-specific safety assessments for patients with active and passive implants.

Cite This Study

Shen et al. (2025) studied Lower limb amputation. Lower limb amputation was evaluated on Specific absorption rate (SAR) and RF-induced lead-tip heating. More extensive lower limb amputations reduced overall specific absorption rate and temperature rise for implants during 1.5 T MRI, but intensified localized heating near the amputation site.

synapsesocial.com/papers/6a930e697a20b505a3a64c4dhttps://doi.org/10.1002/mrm.70123
View Full Paper
Ask AI
Bookmark
Share