Data reduction in non-null tests is difficult due to the presence of retrace error. We propose a simple yet effective data reduction approach for aspheric testing in a non-null interferometer. The new approach gives figure error of the aspheric by just subtracting the theoretical wavefront and first-order errors from the real wavefront obtained in the non-null interferometer. Precise prediction of the theoretical wavefront can be achieved by accurate calibration of the partial compensation system. The approach can be considered a generalization of the traditional data processing method in null tests, and errors that may affect its accuracy are discussed. A set of experiments have been carried out to demonstrate its validity and feasibility.
No takes yet. Share an insight, caveat, or question.
Tian et al. (2012) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: