Key result
Unipolar electrogram-guided ablation predicted transmural lesions with 94% sensitivity and 100% specificity, achieving equal efficacy to Ablation Index guidance for complete bidirectional block.
Why the study?
Modification of the atrial unipolar electrogram reflects transmural loss of conductivity, but the feasibility of unipolar electrogram morphology-guided radiofrequency ablation to generate contiguous, transmural lesions resulting in conduction block compared to Ablation Index parameters was unproven.
Does unipolar electrogram-guided ablation improve the creation of contiguous lines of conduction block compared to Ablation Index-guided ablation in a swine model?
Does unipolar electrogram-guided ablation improve the creation of contiguous lines of conduction block compared to Ablation Index-guided ablation in a swine model?
Unipolar electrogram-guided ablation is a feasible strategy to create contiguous, transmural lesions with bidirectional block while delivering less energy than standard Ablation Index-guided ablation.
May reduce energy for equivalent block in swine; leaves open clinical translation of unipolar EGM guidance.
Background Modification of the atrial unipolar electrogram (Uni- EGM ) with loss of the near-field negative component in response to radiofrequency ablation reflects transmural loss of tissue conductivity. Objective The present study sought to investigate feasibility of Uni- EGM morphology guided radiofrequency ablation (RFA) in generating contiguous, transmural lesions that result in conduction block. This method was compared with ablation guided by standard Ablation Index parameters. Methods In a beating heart swine model, linear transcaval ablation was performed using an irrigated ablation catheter with standard and high power RFA. First, an optimal Uni- EGM endpoint for predicting a transmural lesion was determined by single RF applications in the smooth and trabeculated portions of the right atrium. We compared termination of RF energy immediately upon observing consistent loss of the negative component (s-wave) on the local unipolar electrogram (U +0 ) versus extending RF delivery for an additional 3 seconds beyond (U +3 ). Next, linear ablation was achieved with either target Ablation Index (AI 400 ) or unipolar electrogram (U +3 ) based techniques. The latter method relied upon Uni- EGM morphology to both direct catheter positioning at sites of contiguous and viable tissue and to titrate RF delivery. Bidirectional block was demonstrated with standard EP pacing maneuvers, high density activation mapping and pathology. Results Extending RF delivery for 3 additional seconds (U +3 ) after consistent loss of the negative component on the Uni- EGM proved an optimal endpoint, predicting a transmural lesion with 94% sensitivity and 100% specificity. It was observed that with increasing distance from site of radiofrequency application, the near field s-wave on the Uni- EGM grows in magnitude while an acute current of injury pattern diminishes. Uni- EGM based technique resulted in equal efficacy in producing ablation lines with complete bidirectional block, compared with target AI guided ablation. With standard power, Uni- EGM guided ablation resulted in significantly less energy delivered per RF application. Conclusions We demonstrate that Uni- EGM can successfully guide ablation yielding conduction lines of block. Uni- EGM morphology reflects local tissue activation and its modification in response to thermal injury can serve as an endpoint to ensure adequate yet avoid excess RF delivery. The near-field s-wave on the Uni- EGM indicates viable tissue, and in conjunction with an acute current of injury identified contiguous, conductive tissue for targeting. Characterization of unipolar electrogram morphology with respect to myocardial electrical properties and understanding its modification in response to injury may lend to continued pursuit for safer, more effective ablation strategies.
No takes yet. Share an insight, caveat, or question.
Yavin et al. (2025) studied this question. Unipolar electrogram (Uni-EGM) guided radiofrequency ablation vs. Target Ablation Index (AI 400) guided ablation was evaluated on Transmural lesion prediction and complete bidirectional block. Unipolar electrogram-guided ablation predicted transmural lesions with 94% sensitivity and 100% specificity, achieving equal efficacy to Ablation Index guidance for complete bidirectional block.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: