Multi-element photoacoustic tomography systems have either a non-movable focus or a non-uniform light field distribution over the region of interest, significantly limiting the depth-of-field and resolution of images. In this study, focus-tunable photoacoustic tomography (FT-PAT) based on a clinical ultrasound array transducer with uniform laser excitation was developed to achieve high-resolution imaging in deep tissue. An adjustable line-focusing structure was used to condense the laser beam, and the excited photoacoustic signal was processed via a synthetic-aperture focusing technique. Carbon rods at different depths were used to evaluate the focus-tunable ability in a tissue phantom. In vivo focus-tunable imaging performance was demonstrated by an experiment involving imaging of carbon nanoparticle-labeled living nude mouse covered by chicken breast. The signal-to-noise ratio of targets at different depths was changed by focus adjustment, which demonstrates that the FT-PAT system has potential clinical applications in deep tumor imaging and subsequent diagnoses.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2018) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: