The technique of Nuclear Quadrupole Resonance thermometry is described. It is shown that the N.Q.R. thermometer exhibits unique properties of sensitivity, range, and reproducibility that are all desirable qualities for a temperature standard. The response to temperature is a unique characteristic of the working substance, which in this case is KClO 3 . Experimentally, the center frequency of the absorption line of Cl 35 in pure samples of KClO 3 at 77 °K is reproducible to approximately 2 Hz. This corresponds to an absolute accuracy of better than ± 0.001 °K. The purity of the samples is achieved by recrystallization of commercial reagents. The range of the thermometer extends from 10 °K to approximately 450 °K. A description of a simple system which locks an oscillator to the N.Q.R. absorption line is given. A new method of determining accurately the frequency of the center of the line using magnetic field broadening is also described. With this technique a sensitivity of ± 0.0002 °K is expected. In line with these results, evaluation of a thermometer based on the temperature dependence of the nuclear magnetic resonance frequency in antiferromagnetic salts is made.