A noncontact optical system for generation and detection of ultrasound is described. A confocal Fabry-Perot interferometer combined with a multimode optical fiber is used to reconstruct ultrasound images on a sample with a rough surface. The optical fiber is automatically adjusted to optimal positions for effective collection of scattered light from the sample. The confocal Fabry-Perot interferometer provides sensitive detection independent of surface roughness. The system is applied to a fatigue test sample, with which a conventional interferometer has poor sensitivity because of irregularities on the sample surface. The 2D profiles of the fatigue crack are successfully obtained at temperatures up to 1000°C. The crack length estimated from the profile agrees well with that measured using an optical method.
No takes yet. Share an insight, caveat, or question.
Nakano et al. (1993) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: