Key result
A flexible MEMS sensor measured a mean experimental intravascular shear stress of 11.9 dyn x cm(-2) in the rabbit abdominal aorta, compared to 31.1 dyn x cm(-2) estimated by CFD simulations.
Why the study?
Does a flexible MEMS sensor accurately measure real-time intravascular shear stress in the abdominal aorta of rabbits compared to computational fluid dynamics simulations?
Does a flexible MEMS sensor accurately measure real-time intravascular shear stress in the abdominal aorta of rabbits compared to computational fluid dynamics simulations?
A flexible coaxial-wire-based MEMS sensor can assess real-time intravascular shear stress in the abdominal aorta of rabbits, showing waveform similarity to computational fluid dynamics simulations.
Discrepancy in rabbit aorta leaves MEMS sensor accuracy open; further validation needed before clinical translation.
Fluid shear stress is intimately linked with the biological activities of vascular cells. A flexible microelectromechanical system (MEMS) sensor was developed to assess spatial- and temporal-varying components of intravascular shear stress (ISS) in the abdominal aorta of adult New Zealand white (NZW) rabbits. Real-time ISS (ISS (real-time)) was analyzed in comparison with computational fluid dynamics (CFD) simulations for wall shear stress (WSS). Three-dimensional abdominal arterial geometry and mesh were created using the GAMBIT software. Simulation of arterial flow profiles was established by FLUENT. The Navier-Stokes equations were solved for non-Newtonian blood flow. The coaxial-wire-based MEMS sensor was deployed into the abdominal arteries of rabbits via a femoral artery cutdown. Based on the CFD analysis, the entrance length of the sensor on the coaxial wire (0.4 mm in diameter) was less than 10 mm. Three-dimensional fluoroscope and contrast dye allowed for visualization of the positions of the sensor and ratios of vessel to coaxial wire diameters. Doppler ultrasound provided the velocity profiles for the CFD boundary conditions. If the coaxial wire were positioned at the center of vessel, the CFD analysis revealed a mean ISS value of 31.1 with a systolic peak at 102.8 dyn x cm(-2). The mean WSS was computed to be 10.1 dyn x cm(-2) with a systolic peak at 33.2 dyn x cm(-2), and the introduction of coaxial wire increased the mean WSS by 5.4 dyn x cm(-2) and systolic peak by 18.0 dyn x cm(-2). Experimentally, the mean ISS was 11.9 dyn x cm(-2) with a systolic peak at 47.0 dyn x cm(-2). The waveform of experimental ISS was similar to that of CFD solution with a 30.2% difference in mean and 8.9% in peak systolic shear stress. Despite the difference between CD and experimental results, the flexible coaxial-wire-based MEMS sensors provided a possibility to assess real-time ISS in the abdominal aorta of NZW rabbits.
No takes yet. Share an insight, caveat, or question.
Ai et al. (2009) studied this question. Flexible microelectromechanical system (MEMS) sensor vs. Computational fluid dynamics (CFD) simulations was evaluated on Mean intravascular shear stress (ISS). A flexible MEMS sensor measured a mean experimental intravascular shear stress of 11.9 dyn x cm(-2) in the rabbit abdominal aorta, compared to 31.1 dyn x cm(-2) estimated by CFD simulations.