Key result
A 5-minute cold stress significantly increased maximum radial wall elastance from 0.441 to 0.611 × 10^6 dyne/cm (p = 0.001) compared to room temperature, validating the novel vibration-based measurement method.
Why the study?
Can a sinusoidal minute vibration method quantify dynamic elastance of the superficial arterial wall under different thermal conditions in healthy subjects?
Cross-Sectional (n=29)
No
Can a sinusoidal minute vibration method quantify dynamic elastance of the superficial arterial wall under different thermal conditions in healthy subjects?
Absolute Event Rate: 0.611% vs 0.441%
p-value: p=0.001
The sinusoidal minute vibration method can quantitatively assess superficial arterial wall elastance and its dynamic changes in response to thermal stress, offering a potential tool for evaluating arterial stiffness.
Enables dynamic arterial elastance assessment; leaves open clinical validation before any practice change.
BACKGROUND: The purpose of this study is to propose a novel method for assessing dynamic elastance of the superficial arterial wall using the sinusoidal minute vibration method. METHODS: A sinusoidal signal was used to drive a vibrator which induced a displacement of 0.15 mm with a frequency range between 40 and 85 Hz. The vibrator closely contacted with the wall of a superficial radial artery, and caused the arterial wall to shift simultaneously. A force sensor attached to the tip of the vibrator was used to pick up the reactive force exerted by the radial arterial wall. According to the Voigt and Maxwell models, a linear relationship was found between the maximum reactive force and the squared angular frequency of the vibration. The intercept of the linear function represents the arterial wall elastance. In order to validate the feasibility of our method, twenty-nine healthy subjects were recruited and the wall elastances of their radial arteries were measured at room temperature (25 °C), after a 5-min cold stress (4 °C) and a 5-min hot stress (42 °C), respectively. RESULTS: After the 5-min cold stimulation, the maximum radial wall elastance significantly increased from 0.441 ± 0.182 × 10(6) dyne/cm to 0.611 ± 0.251 × 10(6) dyne/cm (p = 0.001). In the 5-min hot stress, the maximum radial wall elastance significantly decreased to 0.363 ± 0.106 × 10(6) dyne/cm (p = 0.013). CONCLUSIONS: The sinusoidal minute vibration method proposed can be employed to obtain the quantitative elastance of a superficial artery under different thermal conditions, and to help assess the severity of arterial stiffness in conduit arteries.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2016) conducted a cross-sectional in Healthy (n=29). 5-minute cold stress (4 °C) vs. Room temperature (25 °C) was evaluated on Maximum radial wall elastance (10^6 dyne/cm) (p=0.001). A 5-minute cold stress significantly increased maximum radial wall elastance from 0.441 to 0.611 × 10^6 dyne/cm (p = 0.001) compared to room temperature, validating the novel vibration-based measurement method.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: