An iodine-coated polycrystalline platinum electrode was applied to analysis of iron in dietary supplements. The developed method is based on scanning the potential of the iodine-coated polycrystalline platinum electrode between the hydrogen evolution limit (-0.25 V) and a potential lower than the threshold of iodine desorption from the platinum surface (ca. +0.80 V). An anodic peak centered nearly at 0.56 V is assigned to Fe(II) oxidation to Fe(III). The anodic peak current (i p ) showed an excellent linearity (R 2 = 0.995) with the concentration of Fe 2+ within a concentration range from 4-500 ppm Fe. The anodic peak current, as expected for a diffusion controlled process showed a linear relationship with the square root of the scan rate (R 2 =0.999). The limit of detection (based on S/N = 3) for the developed method is 2.34 ppm while the limit of quantitation (based on S/N = 10) is 7.8 ppm. The effect of potential interferences like calcium, magnesium, zinc, selenium, copper, manganese, chromium, molybdenum, vanadium was investigated. Silver and copper were the only ones among the investigated elements which showed resolved peaks from those of Fe (II) within the accessible potential window of the iodine-coated platinum electrode. Analysis of real samples and application of statistical tests to the results indicated the absence of systematic error in the results and the validity of the null hypothesis, i.e., there is no significant difference between the nominal values reported by the manufacturers and our results at p=0.05.