Key result
Dual plane fluoroscopic imaging with automatic 2D-3D image-model registration provided highly repeatable in vivo shoulder kinematics (0.50mm translation, 1.04° rotation) compared to single plane.
Dual plane fluoroscopic imaging with automatic 2D-3D registration provides highly accurate and repeatable measurements of shoulder joint kinematics compared to single plane techniques.
Answers are grounded in this paper and cite their sources.
May refine shoulder kinematics assessment in research; leaves open clinical validation before practice adoption.
Fluoroscopic imaging, using single plane or dual plane images, has grown in popularity to measure dynamic in vivo human shoulder joint kinematics. However, no study has quantified the difference in spatial positional accuracy between single and dual plane image-model registration applied to the shoulder joint. In this paper, an automatic 2D-3D image-model registration technique was validated for accuracy and repeatability with single and dual plane fluoroscopic images. Accuracy was assessed in a cadaver model, kinematics found using the automatic registration technique were compared to those found using radiostereometric analysis. The in vivo repeatability of the automatic registration technique was assessed during the dynamic abduction motion of four human subjects. The in vitro data indicated that the error in spatial positional accuracy of the humerus and the scapula was less than 0.30mm in translation and less than 0.58° in rotation using dual plane images. Single plane accuracy was satisfactory for in-plane motion variables, but out-of-plane motion variables on average were approximately 8 times less accurate. The in vivo test indicated that the repeatability of the automatic 2D-3D image-model registration was 0.50mm in translation and 1.04° in rotation using dual images. For a single plane technique, the repeatability was 3.31mm in translation and 2.46° in rotation for measuring shoulder joint kinematics. The data demonstrate that accurate and repeatable shoulder joint kinematics can be obtained using dual plane fluoroscopic images with an automatic 2D-3D image-model registration technique; and that out-of-plane motion variables are less accurate than in-plane motion variables using a single plane technique.
No takes yet. Share an insight, caveat, or question.
Zhu et al. (2012) studied Shoulder joint kinematics measurement (n=4). Dual plane fluoroscopic images with automatic 2D-3D image-model registration vs. Single plane fluoroscopic images was evaluated on In vivo repeatability in translation and rotation. Dual plane fluoroscopic imaging with automatic 2D-3D image-model registration provided highly repeatable in vivo shoulder kinematics (0.50mm translation, 1.04° rotation) compared to single plane.
Synapse has enriched 10 related papers. Consider them for comparative context: