A simulant with a suitable surface and the correct thermal inertia, β=kρc, meets the primary requirement of accepting heat at the same rate as the skin. For protected skin, a simulant with a blackened opaque surface, β½=0.035 cal·cm−2·sec−½·°C−1, and equivalent depth x/2√α=0.12 sec12 (α=k/ρc) is suitable. Such simulants have been made of fused silica and of an epoxy resin filled with aluminum powder. The detector was a butt welded copper-constantan thermocouple, flattened to a thickness of 0.0015 cm, or a thermocouple of evaporated gold and bismuth films, ∼6000 Å thick. Results with water flow simulating blood flow in the skin indicate that an inert system simulates the response of skin rather well; if some adjustment must be made for varying blood flow it can be done experimentally or by calculation.
No takes yet. Share an insight, caveat, or question.
John Davies (1970) studied this question.