We describe an instrument for measuring linear retardance in transparent optical components using two photoelastic modulators. The instrument contains a He-Ne laser (632.8 nm), a polarizer, two low- birefringence photoelastic modulators at different frequencies, an ana- lyzer, and a silicon photodiode detector. A sample is placed between the two modulators. The detector signals corresponding to linear retardance in a sample are analyzed using lock-in amplifiers. A computer program calculates and displays both the retardation magnitude and the angle of the fast axis. The reported instrument is essentially a polarimeter specifi- cally designed for measuring low-level linear retardance in high-quality optical components. It provides a retardation sensitivity of better than 0.005 nm (0.003 deg or 5310 25 rad with a He-Ne laser at 632.8 nm).
No takes yet. Share an insight, caveat, or question.
Baoliang Wang (2002) studied this question.