A VG9000 glow discharge mass spectrometer has been modified for the direct analysis of solid nuclear samples within a glove-box environment. Because containment is needed for the analysis of this kind of material, the glove-box encloses all parts of the instrument that come into contact with the sample, namely the ion source chamber, sample interlock and associated pumping system. External modifications eliminate outside contamination by the fitting of absolute filters on all source supplies. Internally, the design of the ion source has been altered to minimize the number of operations performed inside the glove-box thereby simplifing operation and routine maintenance. These modifications retain the ion extraction and focusing properties of the instrument. The data presented show that there is no compromise in the analytical performance of the instrument when placed in the glove-box. Data representative of nuclear materials is also shown.