A cylindrical 16-pole electromagnet (EM) for electron paramagnetic resonance imaging (EPRI) and low-field magnetic resonance imaging (MRI) has been designed by means of two-dimensional and three-dimensional (3-D) finite element analysis (FEA). The use of an automatic procedure that combines FEA with a minimization routine allowed the optimization of the design, in order to improve the homogeneity along the axis of the EM. A prototype has been built by using electrical steel sheets that were cut by laser; this solution reduced significantly the manufacturing cost. The EM operates with a maximum flux density, in the bore, of 0.08 T and has a homogeneity along the axis of about 40 parts per million (ppm) in a spherical region 10 cm in diameter. It generates the main field and two of the three field gradients required in the 3-D image reconstruction. Good agreement was found between the results of simulation and the measured values.
No takes yet. Share an insight, caveat, or question.
Chiricozzi et al. (1998) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: