A commercial scanning acoustic microscope operating in the frequency range of 50–2000 MHz is described. It has a stable scan mechanism suitable for the high-frequency end to give high-resolution images and modes to allow for material characterization at the low-frequency end. The high-frequency electronics permitting a wideband operation with a variety of acoustic lenses are described. Microprocessors control many parts of the the instrument to release the user from the routine adjustments, resulting in a user-friendly interface. Images are presented to show the resolution and penetration abilities of the instrument at several frequencies, along with comparative optical images.
No takes yet. Share an insight, caveat, or question.
Atalar et al. (1986) studied this question.