RATIONALE AND OBJECTIVES: The detection of arthritis by autofluorescence was investigated using an antigen-induced arthritis model. METHODS: For autofluorescence investigations of joints, a mobile fluorescence-detector was constructed consisting of a lens/mirror system attached to a conventional spectrofluorometer and optimized fiber optic cables reaching to and from the site of investigation. Autofluorescence measurements were performed at 7 arthritic and 7 healthy mice. Fifteen antigen-induced arthritis and 3 healthy mice were used for histologic examinations. RESULTS: In the exudative stage (day 1), a decrease of emission signal intensities for excitation wavelengths at 300 nm (emission, 355-365 nm) and 360 nm (emission, 475-485 nm) was observed. Signals increased on day 7 (maximum of cellular infiltration). Chronic inflammation (day 14 and 21) led to a decrease of signals again. CONCLUSION: Arthritis influences autofluorescence signals in vivo. The detected excitation/emission pairs can be assigned to collagen/elastin and NAD(P)H. Signal intensities of NAD(P)H differed significantly from controls at day 1 and 7.
No takes yet. Share an insight, caveat, or question.
Hansch et al. (2003) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: