Coated quartz crystal microbalances were modified with a surface-imprinting process using whole yeast cells. These molded polymer and sol-gel surfaces show honeycomb-like structures as shown by atomic force microscopy. Reinclusion of cells allows a selective on-line monitoring of these microorganism concentrations in water over 5 orders of magnitude. The sensitivity to cells holds up in growth media up to 21 g/L. Even cell fragments can be detected in flowing conditions. The highly robust polymers on the sensor devices are suitable for biotechnological applications.
No takes yet. Share an insight, caveat, or question.
Dickert et al. (2002) studied this question.