OBJECTIVES: The aim of this pilot study was to compare the image quality of a storage phosphor system with that of conventional film-screen in pre-implant conventional tomography, and to test the influence of radiation dose on image quality in the storage phosphor system. METHODS: Cross-sectional conventional tomographic images (Scanora) technique) were obtained on 11 patients with film-screen and with storage phosphor image plates (Digora) PCT) at normal and low doses. Ten observers graded the visibility of anatomical structures of importance for implant planning. A three-step rating scale was used, where -1 =worse, 0=equal and 1=better than the reference image. RESULTS: Although image quality was graded as equally good in the majority (59%) of images, the storage phosphor system scored significantly lower than film-screen (-0.37 vs 0.00, respectively) for all the images. Low dose storage phosphor images were rated significantly lower (-0.21) than normal dose images (0.00). CONCLUSIONS: In the majority of patients, anatomic structures of importance for implant planning are visualized equally well on storage phosphor and film-screen images. However, where differences do exist, storage phosphor images score lower than film-screen images. Dose reduction in the storage phosphor system had a negative influence on image quality.