We investigated the possibility of significant corneal trauma (as revealed by slit lamp observation of the fluorescein instilled eye), and massage effects following determination of intraocular pressure with the A.O. Non-Contact tonometer (NCT). Fifteen different, normal human eyes were each applanated 150 successive times with the NCT; leading to the conclusion that only minor, superficial corneal epithelial defects sometimes resulted and that, in line with other studies, the initially higher readings (about 1 mm), obtained with the NCT, were most likely due to patient apprehension, while the subsequently lower readings represented patient acceptance of the process and were not a result of true aqueous massage. As in an earlier study, we found the instrument's variability to be about ±1 or ±2 mm and probably due to the subject's own cardiac cycle.