Answers are grounded in this paper and cite their sources.
The objective of this study was to evaluate the fixation strength of a biodegradable interference screw (Arthrex, Naples, FL) compared with press fit fixation and a titanium interference screw in anterior cruciate ligament (ACL) reconstruction using a bone-patellar tendon-bone (BPTB) graft. Porcine lower limbs were used. Ultimate failure loads of the biodegradable screw (805.2 N; range 680 to 995 N) did not differ significantly from ultimate failure loads of titanium interference screws (768.6 N; range 544 to 1094 N). Press fit fixation provided significant (P < .01) lower fixation strength (462.5 N; range 80 to 825 N). These results support the conclusion that biodegradable interference screws are a reasonable alternative in terms of primary fixation strength.
No takes yet. Share an insight, caveat, or question.
Rupp et al. (1997) studied this question.