An evaluation of flow-injection analysis with chemiluminescence detection (FIA-CL) to quantify Fe(2+) ((aq)) in freshwaters was performed. Iron-coordinating and/or iron-reducing compounds, dissolved organic matter (DOM), and samples from two natural water systems were used to amend standard solutions of Fe(2+) ((aq)). Slopes of the response curves from ferrous iron standards (1 - 100 nM) were compared to the response curves of iron standards containing the amendments. Results suggest that FIA-CL is not suitable for systems containing ascorbate, hydroxylamine, cysteine or DOM. Little or no change in sensitivity occurred in solutions of oxalate and glycine or in natural waters with little organic matter.