REMEDi (Rapid EMErgency Drug identification; Bio-Rad) is an automated high performance liquid chromatographic (HPLC) system designed to detect, identify and measure a range of basic and neutral drugs in 0.5-1.0 mL of urine or plasma/serum. We have evaluated REMEDi in the analysis of the antiarrhythmic drug disopyramide in patient samples. The specimens were also analysed by a conventional HPLC method, based on solvent extraction and UV detection (254 nm), and by EMIT. There were good correlations between the results obtained with each method (r = 0.91 or greater). REMEDi gave a lower mean result than EMIT [means +/- SD (mg/L): REMEDi 2.64 +/- 1.10, EMIT 3.14 +/- 1.51; t = 4.0, p less than 0.01; n = 25], but there were no other significant differences in mean results. The principal disopyramide metabolite, mono-N-desalkyldisopyramide, did not interfere in any method. Clearly REMEDi can be used for therapeutic drug monitoring of disopyramide provided enough sample is available.
No takes yet. Share an insight, caveat, or question.
Patel et al. (1991) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: