Answers are grounded in this paper and cite their sources.
Intravascular ultrasound (IVUS) plays an important role for the detection of arteriosclerosis, which causes the ischemic heart disease. In mechanical scanning-type IVUS, it is necessary to rotate a transducer or a reflecting mirror. A method that involves rotating the transducer using a torque wire causes image distortion (NURD: non uniform rotation distortion). For a method that involves placing an electromagnetic motor on the tip of an IVUS probe is difficult to miniaturize the probe. Our objectives are to miniaturize the probe (1 mm in diameter, 5 mm in length) and to remove NURD. Therefore, we conducted a study to assess the feasibility of attaining these objectives by constructing a prototype IVUS system, in which an ultrasound motor using a stator in the form of a helical coil (abbreviated as CS–USM: coiled stator–ultrasonic motor) is incorporated, and to clarify problems that need to be solved in constructing the probe.
No takes yet. Share an insight, caveat, or question.
Tanabe et al. (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: