Answers are grounded in this paper and cite their sources.
A hypoxanthine sensor was constructed using a polarographic electrode in combination with immobilized xanthine oxidase [E.C. 1.2.3.2.]. Electrodes for the determination of both the generated hydrogen peroxide and the consumed oxygen were tested. Xanthine oxidase was immobilized via glutaraldehyde activation on a polycarbonate membrane which was subsequently mounted on the tip of the electrodes. A linear response of the sensor in the range 2.5 – 375 × 10−6 M hypoxanthine was observed. An inosine sensor containing xanthine oxidase and nucleoside phosphorylase [E.C. 2.4.2.1.] was also developed using the same principles. This sensor responded linearly to inosine concentrations in the range 2.5 – 500 × 10−6 M. The determination of hypoxanthine and inosine concentrations is of importance in fish freshness estimations.
No takes yet. Share an insight, caveat, or question.
Haemmerli et al. (1990) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: