Key result
The robotic ICE catheter manipulator achieved an average in vivo catheter tip localization error of 1.5 mm and an average image similarity of 81.1% during automated view recovery.
Why the study?
Reliance on intraprocedural intra-cardiac echocardiography places high cognitive demand on physicians who often act as both interventionalist and imager.
Answers are grounded in this paper and cite their sources.
A novel robotic manipulator for ICE catheters featuring automated view recovery and kinematic compensation shows promise in improving catheter control and imaging efficiency in preclinical models.
Intra-cardiac Echocardiography (ICE) is a powerful imaging modality for guiding electrophysiology and structural heart interventions. ICE provides real-time observation of anatomy, catheters, and emergent complications. However, this increased reliance on intraprocedural imaging creates a high cognitive demand on physicians who can often serve as interventionalist and imager. We present a robotic manipulator for ICE catheters to assist physicians with imaging and serve as a platform for developing processes for procedural automation. Herein, we introduce two application modules towards these goals: (1) a view recovery process that allows physicians to save views during intervention and automatically return with the push of a button and (2) a data-driven approach to compensate kinematic model errors that result from non-linear behaviors in catheter bending, providing more precise control of the catheter tip. View recovery is validated by repeated catheter positioning in cardiac phantom and animal experiments with position- and image-based analysis. We present a simplified calibration approach for error compensation and verify with complex rotation of the catheter in benchtop and phantom experiments under varying realistic curvature conditions. Results support that a robotic manipulator for ICE can provide an efficient and reproducible tool, potentially reducing execution time and promoting greater utilization of ICE imaging.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2020) studied Intra-cardiac Echocardiography (ICE) manipulation (n=3). Robotic ICE catheter manipulator vs. Without compensation / Manual manipulation was evaluated on In vivo catheter tip localization error during automated view recovery (1.5 mm). The robotic ICE catheter manipulator achieved an average in vivo catheter tip localization error of 1.5 mm and an average image similarity of 81.1% during automated view recovery.