Techniques for improving the relative accuracy of longwave radiation measurements by a set of pyrgeometers [the Eppley Laboratory Precision Infrared Radiometer (Model PIR)] are presented using 10 PIRs from the 1999 Cooperative Atmosphere–Surface Exchange Study (CASES-99). The least squares–based optimization technique uses a field intercomparison (i.e., a time period during which all the PIRs were upward looking and set up side by side) to determine a set of optimization coefficients for each PIR. For the 10 CASES-99 PIRs, the optimization technique improved the standard deviation of the difference of downwelling irradiance between the PIRs from 60.75 to 60.4 W m22 (for nighttime data). In addition to presenting the optimization method, various PIR data quality checks are outlined and applied to the PIR data. Based on these quality checks, the measured case and