Objective The main aim of this study was to evaluate the usefulness of urine reagent strip tests in the diagnosis of meningitis. Patients and methods Fifty children with a mean age of 6.53+4.55 years who were clinically suspected to have meningitis were recruited from Benha University Hospital and Shebin El Koom Fever Hospital. Cerebrospinal fluid (CSF) samples were tested using both Combur-10 urinary reagent strip and CSF microscopy examination as reference standards. Results The diagnostic accuracy of protein estimation by reagent strip showed 100% sensitivity and 3.7% specificity for the detection of CSF proteins at the level of 100 mg/dl, whereas the specificity was 50% at the level of 47 mg/dl. Glucose reagent strip had 96.8% sensitivity and 100% specificity for the detection of CSF glucose at the level of 46 mg/dl. Leukocyte esterase positivity by test strip had a sensitivity of 100% and a specificity of 75% for the detection of CSF granulocytes of more than 10 granulocytes/mm. There was a significant increase in CSF parameters (leukocyte, protein) by both methods and a highly significant increase in glucose in the septic group compared with the aseptic group. Conclusion Urinary reagent strips offer a clinically useful and rapid screening method for the quantitative determination of CSF, glucose, protein, and leukocytes.