Answers are grounded in this paper and cite their sources.
Various straight-jawed aneurysm clips were tested for occluding capabilities on a vascular tissue model. Occluding pressures varied markedly among the clip styles and were altered by changes in the lumen and tissue composition of the model. Mechanical characteristics are highly variable between clip styles, but fall within a narrow range for clips of a similar style. The complex interplay of the unique aspects of clip design and force generated by the spring in conjunction with tissue characteristics and precise clip placement upon the tissue are major variables in the apparent ability of a particular clip to occlude the neck of an aneurysm.
No takes yet. Share an insight, caveat, or question.
Rosenbaum et al. (1978) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: