A field portable gas chromatograph (GC) was constructed allowing the enrichment of organic solutes from air samples on a miniaturized chemical trap and the subsequent gas chromatographic analysis on a resistively heated capillary column. The heart of the system is an integrated chip-based plasma emission detector (muPED). As a non-selective detector, the sensitivity is similar to that of a flame ionization detector (FID). The detector shows good selectivity for phosphorus, sulfur and chlorine-containing compounds with relative selectivities of ca. 5 x 10(5) gP gC(-1), 50 gS gC(-1) and 10(2) gCl gC(-1). The lifetime of the plasma chip under air monitoring conditions exceeded 3000 analyses.