Background: Massive osteolysis of the proximal femur makes stem revision a challenging procedure. EBRA-FCA provides the opportunity to determine stem migration, which is considered a predictive factor for implant survival. In this study, we aimed to analyze the migration behavior of a modular, distally fixed reconstruction prosthesis. Methods: Applying a retrospective study design, we reviewed all consecutive patients who received a cementless MP reconstruction prosthesis (Waldemar Link GmbH & Co. KG, Hamburg, Germany) at our Department between 2005 and 2019. We reviewed medical histories and performed radiological measurements using EBRA-FCA software. Results: A total of 67 stems in 62 patients (female 26; male 36) fulfilled our inclusion criteria. Mean age at surgery was 68.0 (range 38.7−88.44) years. EBRA migration analysis showed a median subsidence of 1.6 mm (range 0.0−20.6) at 24 months. The angle between stem and femur axis was 0.3° (range 0.0°−2.9°) at final follow-up. No correlation between body mass index and increased subsidence was found (p > 0.05). Overall revision-free rate amounted to 92.5% and revision-free rate for aseptic loosening to 98.5%. Furthermore, no case of material breakage was detected. Conclusions: In summary, the MP reconstruction prosthesis showed low subsidence and reduction in the migration rate over the investigated follow-up. Based on this, the modular stem can be considered as a good therapy option in challenging stem revisions offering various options to address the individual anatomical situation.
No takes yet. Share an insight, caveat, or question.
Blum et al. (2022) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: