Answers are grounded in this paper and cite their sources.
PURPOSE: Dislocation after total hip arthroplasty (THA) is a common reason for revision. The last decade fostered a significant increase in the use of dual-mobility cups (DMCs). Here we report our study on the short-term survival rate of a cemented DMC reported to the Swedish Hip Arthroplasty Register (SHAR) compared with other cemented designs used in first-time revision due to dislocation. METHODS: During 2005-2015, 984 first-time revisions for dislocation were reported to SHAR. In 436 of these cases a cemented dual articular cup was used. During the same time period, 355 revisions performed with a standard cemented cup (femoral head size 28-36 mm) were reported to the SHAR. Patients receiving a DMC were slightly older (75 years, p = 0.005). Re-revision for all reasons was used as primary endpoint. We also anlaysed risk for re-revision of the acetabular component and re-revision due to dislocation. Kaplan-Meier implant survival and a Cox regression analyses adjusted for age and gender were performed. RESULTS: Implant survival at 4 years for all reasons (91% ± 3.7% vs 86% ± 4.1%, p = 0.02), and especially for re-operation because of dislocation, favours the DMC group (96% ± 3.0% vs 92% ± 3.3%, p = 0.001). DISCUSSION: Our findings indicate that use of a cemented DMC reduces the short- to mid-term risk of a second revision in first-time revisions compared with classic cup designs. Longer follow-up is needed to establish any long-term clinical advantages when DMCs are used in revisions performed due to dislocation.
No takes yet. Share an insight, caveat, or question.
Mohaddes et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: