20 patients, who had had total hip replacements for symptomatic osteoarthrosis secondary to Paget's disease, were followed for a mean of 6 (4-8) years. Proximal hydroxyapatite-coated stems were implanted in all patients. 12 patients received hydroxyapatite-coated, 2 cemented (Muller type) and 6 cementless cups (Morsher type). The mean Harris hip score was 31 (7-40) points preoperatively and 88 (74-100) postoperatively. The radiographic evaluation revealed good stability and fixation using Engh's criteria. One stem subsided early more than 5 mm and then seemed to stabilize. Our findings support the use of hydroxyapatite total hip implants for patients with this disease and osteoarthrosis.
No takes yet. Share an insight, caveat, or question.
Kirsh et al. (2001) studied this question.