CASES: Three patients (61-81 years) experienced abrupt shoulder pain and functional decline more than 5 months after reverse total shoulder arthroplasty (RTSA) using a monoblock central-screw baseplate in graft-reconstructed glenoids. Imaging revealed central screw breakage with variable baseplate displacement. Two underwent successful revision with restoration of stable fixation; in the third, revision was deferred for medical comorbidities. CONCLUSION: Central screw breakage is an important marker of glenoid fixation failure after RTSA. These cases suggest an association with graft-dependent baseplate fixation, limited native bone support, and consequently late fatigue failure. Prompt diagnosis and timely management are essential.
No takes yet. Share an insight, caveat, or question.
Romem et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: