The instantaneous frequency of a rapidly tuned continuous-wave (CW) laser is measured through linear optical sampling against dual-frequency combs. This dual-comb interferometer determines the instantaneous frequency of the CW laser during a quasi-sinewave frequency sweep of 3 THz amplitude with a 10 ms period. More complicated waveforms are also measured with instantaneous chirps exceeding 1500 THz/s (12 000 nm/s). The uncertainty is 1.5 MHz at 20 ns time resolution, averaging down to 5 kHz at 5 μs time resolution. The absolute frequency accuracy can be calibrated to within 2.5 kHz provided there is a brief period (<;1 ms) of low laser chirp (<;160 GHz/s) during the waveform measurement to allow for a dual-comb Vernier measurement of the absolute frequency, modulo 3 THz. This approach allows for the characterization of arbitrary CW waveforms with instantaneous frequencies that change rapidly and over wide optical bandwidth.
No takes yet. Share an insight, caveat, or question.
Coddington et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: